Filter Tester

FNM's Filter Testers, including the FT150PA, FT200PO, and FT200PS models, are designed to assess the efficiency and pressure drop characteristics of various flat sheet filter media. These testers utilize oil/salt aerosols or atmospheric dust to conduct the assessments. Our filter testers are versatile tools suitable for evaluating a range of flat sheet media commonly employed in the production of respiratory face masks and air filters.

Filter testers are equipped with high precision particle counter (six-channel particle sizes: 0.3, 0.5, 1.0, 2.5, 5.0 and 10.0 µm) for counting the particles generated by oil / salt particle generator or atmospheric dust in order to calculate filtration efficiency. Additionally, this machine can calculate air permeability and pressure drop of flat sheet media automatically.
FT200PO and FT200PS use two particle counters that simultaneously count upstream and downstream particles. Data can be saved automatically and also test reports can be printed by an installed printer on the machine.
FNM’s filter testers can measure the efficiency and pressure drop of flat sheet media according to BS EN 143, BS EN 149, BS EN 779, ISO 16890, ISO 16900-3, ISO 11155-1 and ISO 5011 standards.
Note: FNM’s filter testers are designed for testing flat sheet air filter media only.

 

FT150EA

FT200PO

FT200PS

Test Modes

Pressure drop test

Air permeability test

Filtration efficiency test

Atmospheric

Atmospheric and oil aerosols

Atmospheric, oil and salt aerosols

Bubble point

Optional

Optional

Optional

Standards

BS EN 149

BS EN 779

US 42 CFR 84

ISO 16890

ISO 16900-3

ISO 11155-1

ISO 5011

Control

PLC

HMI

7”

7”

7”

Air flow

Flow

10 - 150 l/min

10 - 200 l/min

10 - 200 l/min

Digital control

Media Holder

Test area

4.9, 20, 25, 38, 50 and 100 cm2

4.9, 20, 25, 38, 50 and 100 cm2

4.9, 20, 25, 38, 50 and 100 cm2

Sensors (Transmitters)

Temperature

Relative humidity

Digital tank pressure control

Air Compressor

Tank pressure

1 – 8 bar

1 – 8 bar

1 – 8 bar

Pressure Drop

Pressure drop

0 – 760 Pa

0 – 760 Pa

0 – 760 Pa

Digital control

Air Dryer

Air line trap

Dryer system

Optional

Optional

Optional

Particle Counter

Laser particle counter

1

1

1

Channels

four-channel

four-channel

four-channel

Channel sizes

0.3, 0.5, 1.0, 3.0 μm

0.3, 0.5, 1.0, 3.0 μm

0.3, 0.5, 1.0, 3.0 μm

Flow rate

2.8 L/min

2.8 L/min

2.8 L/min

Aerosol Generator (Oil)

Generator

Aerosol Generator (Salt)

Generator

Neutralizer

Aerosol Dilutor

Particle dilution

100:1

100:1

Type of aerosol challenge

Paraffin, PAO, DOP

Paraffin, PAO, DOP

Printer

Optional

Optional

Respiratory Face Mask Holder (3D masks)

Optional

Optional

Power Supply

Single phase, 220 V, AC

 

Single phase, 220 V, AC

 

Single phase, 220 V, AC

 

Weight

About 300 kg

About 310 kg

About 310 kg

Size (cm) (Length × Width × Height)

90×85×160

90×85×160

90×85×160

Downloads

Request a Quote